9.6
Boreal Laser Inc.
GasFinderMC Installation Manual — Appendices
July 17th 2017
Explanation of R
2
(continued)
A typical plot of
concentration
versus
R
2
will give the following graph:
With lower levels of sample gas, the
R
2
s
decrease, and equal zero when there is
no gas present. As the signal from the gas
becomes stronger, the effect of noise, both
electronic and optical, is reduced and the
R
2
s will increase (i.e., the signal to noise
ratio will increase). The general shape of the
plot is the same for all gases; however, the
x-axis values will depend on the sensitivity
of the instrument to the gas species being
observed.
Use of R
2
The point at which the bend or ‘knee’ of the curve occurs in the plot of R
2
/ ppmm
can be used
to determine the sensitivity of the instrument to the gas concentration measurement. This
varies with the instrument and the type of gas but an R
2
of 0.85 is generally considered to be
the point at which the readings can be considered valid. Below this the readings should be
treated with suspicion.
Summary of Contents for GasFinderMC-2
Page 8: ...viii Boreal Laser Inc GasFinderMC Operation Manual Contents July 17th 2017 ...
Page 24: ...3 14 Boreal Laser Inc GasFinderMC Operation Manual Installation July 17th 2017 ...
Page 30: ...4 6 Boreal Laser Inc GasFinderMC Operation Manual Data I O July 17th 2017 ...
Page 71: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm ...
Page 73: ......
Page 75: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......
Page 89: ......