Appendix C — Explanation of ppmm
Gas concentration is measured in ppm, ‘parts per million’ by volume. If a room measuring
100m x 100m x 100m (1 million cubic metres) has 10 cubic metres of air replaced by a pure
gas, then the gas concentration is expressed as 10ppm. A point sensor measures directly in
ppm. Open path monitors, like the
GasFinderMC
, measure the total amount of a specifi c gas,
for example HF, in the path of the laser beam between the transmitter unit and a refl ector.
This is a ‘total path’ measurement. The units are ‘parts per million metres’, or ‘ppmm’.
A uniform background concentration of 10ppm over 50m gives a reading of 500ppmm.
10ppm x 50m = 500ppmm
When the 500ppmm is divided by 50m the result is a value of 10ppm, which is the ‘path
averaged’ concentration.
A concentrated cloud of 50ppm, 10m in diameter, in a background of 0ppm also gives a
reading of 500ppmm. 50ppm x 10m = 500ppmm
50ppm
50ppm
50m
10m
OP3 Remote Head
Retrorefl ector
Retrorefl ector
10ppm
50m
OP3 Remote Head
Summary of Contents for GasFinderMC-2
Page 8: ...viii Boreal Laser Inc GasFinderMC Operation Manual Contents July 17th 2017 ...
Page 24: ...3 14 Boreal Laser Inc GasFinderMC Operation Manual Installation July 17th 2017 ...
Page 30: ...4 6 Boreal Laser Inc GasFinderMC Operation Manual Data I O July 17th 2017 ...
Page 71: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm ...
Page 73: ......
Page 75: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......
Page 89: ......