9.5
Boreal Laser Inc.
GasFinderMC Installation Manual — Appendices
July 17th 2017
Appendix D — Explanation of R
2
When the
GasFinderMC
receives the returning laser signal after it has passed through the
sample gas, the electronics convert it to the shape of a specifi c waveform or curve. This
is the sample waveform. The
GasFinderMC
also has a similar signal from the calibration
waveform. These curves are then digitised and compared as two numeric arrays.
An accepted mathematical procedure to compare curves or numeric arrays is the
Linear
Least Squares Regression
analysis. This analysis results in a measure of the similarity
(
R
2
),
between the waveform of the
sample
gas and that of the
reference
cell gas. A perfect
similarity would give a value for
R
2
of 1.0, and a total mismatch would be 0.0.
The blue line represents the
Linear Least Squares
fi t of the data and is the best fi t of a
straight line between the reference (X) and sample (Y) data points. The slope is a component
in the ratio-metric calculation of gas concentration.
Regression Analysis for
Waveform B
R2 = 0.04
Sample Waveform B
R2 = 0.04
Sample Waveform A
R2 = 0.98
Regression Analysis for
Waveform A
R2 = 0.98
Summary of Contents for GasFinderMC-2
Page 8: ...viii Boreal Laser Inc GasFinderMC Operation Manual Contents July 17th 2017 ...
Page 24: ...3 14 Boreal Laser Inc GasFinderMC Operation Manual Installation July 17th 2017 ...
Page 30: ...4 6 Boreal Laser Inc GasFinderMC Operation Manual Data I O July 17th 2017 ...
Page 71: ...BOREAL LASER INC MC Instrument Overall Dimensions JB0125MC A 1 of 1 NTS A Dim mm ...
Page 73: ......
Page 75: ......
Page 79: ......
Page 81: ......
Page 83: ......
Page 85: ......
Page 87: ......
Page 89: ......