Setup
Hardware Setup
NeuroPort System
1. Remove the NSP, Amplifier and APS from the shipping boxes.
2. Attach the NSP rack-mount brackets to install it in an equipment rack, or
rubber feet to place it on a table.
3. Plug in the NSP and APS to electrical outlets.
4. Connect one end of the APS cable to the APS and the other end to the
Amplifier. The cable needs to click and lock in place.
5. Connect one end of the fiber-optic cable to the NSP and the other end to
the Amplifier.
6. Attach optional headstages to the amplifier inputs and connect them to
electrodes.
7. Turn on the NSP and APS
8. The LED on the APS, Amplifier, and the NSP should turn green.
Software Setup
Installing Central Software Suite
1. Uninstall any previously existing versions of Central Suite through
Programs and Features in Windows Control Panel.
2. Run the Central Suite installer (.msi) from the installation CD or
Blackrockmicro.com.
3. Install the MATLAB Compiler Runtime from the supplied installation CD or
Blackrockmicro.com.
4. Blackrock recommends that a dedicated PC be used to run Central.
Refrain from installing any additional software or connecting the PC to the
internet.
Setting up the Ethernet Connection
Note: If you purchased the NeuroPort Host PC, the Ethernet card is pre-installed
and configured in your PC.
1. Once installed, the card needs to be configured to connect to the NSP.
2. Click on Start and search for “View Network Connections”.
3. Right click on the correct Ethernet Adaptor (usually, Local Area
Connection) and click on Properties.
4. Uncheck all services except for Internet Protocol (TCP/IP) or Internet
Protocol Version 4 (TCP/IPv4).
5. Click on Internet Protocol (TCP/IP) and click on Properties
6. For IP Address enter 192.168.137.1, for Subnet Mask enter
255.255.255.0 and leave the rest blank.
7. Click on OK to save changes.