![Blackrock Microsystems NeuroPort Instructions For Use Manual Download Page 22](http://html1.mh-extra.com/html/blackrock-microsystems/neuroport/neuroport_instructions-for-use-manual_2767812022.webp)
Revision 1.00 / LB-1070 – NeuroPort Biopotential Signal Processing System IFU
© 2021 Blackrock Microsystems, LLC
22
The On/Off switch is located on the back of the unit. Once it is switched on, the APS will
start a power-up sequence of analog and digital supplies ending with the green ON LED
illuminating. When the power switch is turned off this sequence is reversed, and the
output of the APS is tied to ground. In the event of an error in voltage or power delivery,
the APS will shut down and a red error LED will turn on. After checking all Amplifier and
patient cable connections for any inadvertent shorting, turn off the APS and turn it back
on to reset the error.
Cables
In addition to the above hardware components, the system also comes with a few
cables. Note that standard power cables for the NSP and Front End Amp Power Supply
are included, but not shown below. The fiber-optic cable is used to connect the Front
End Amplifier to the Neural Signal Processor, the Ethernet cable is used to connect the
Neural Signal Processor to the recording computer, and the Front End Amplifier Power
Supply cable is used to connect the Front End Amplifier to its power supply. The Digital
NeuroPort System does not include the front end amplifier power supply cable.
Figure 8–Fiber-optic Cable
Figure 9–Crossover Cable
Figure 10–Amplifier Power Supply (APS) Cable