Neural Modulation
Neural Modulation displays a histogram of changes in firing frequency over time. The
Modulation Index is the variance of the histogram entries and covers a range of 0 to
12.0. A range of 2 to 5 is considered normal when recording neural signals.
Figure 39–Neural Modulation