Analog Output
Analog output properties are accessed from the left side bar of the Hardware
Configuration window and allows the user to set the properties of the analog
output ports. Analog outputs can function as monitor channels, directly mirroring
signals of interest, or as waveform generators. Each modality has output and
triggering parameters defined below.
Label
: User-defined channel label.
Function
: Assigns the analog channel function.
Analog Output: Continuous Monitor
An analog output channel configured as a continuous monitor will scale a user
input signal into a plus or minus 5 V output signal.
Input Range
: Input voltage range displayed for reference.
Output Range
: Output voltage range displayed for reference.
Track Most Recently Selected
: When enabled, the channel selected in Spike
Panel, Single Neural Channel, or Raster Plot view is used as the monitor input.
Channel
: An analog output channel configured as a spike monitor will scale a
user selected input signal into a plus or minus 5 V output signal.
Analog Output: Spike Monitor
An analog output channel configured as a spike monitor will scale a user
selected input signal into a plus or minus 5 V output signal containing only
identified spike events.
Input Range
: Input voltage range displayed for reference.
Output Range
: Output voltage range displayed for reference.
Figure 19–Analog Output: Continuous Monitor