Bird Site Analyzer
26
well as on the mode configure screens. The table below
lists typical worst case values for selected measure-
ments.
1. Turn on the instrument and enter the Measure
Match mode (refer to Select the Measure Match
Mode, page 19).
2. Select Frequency (Setting the Measurement Fre-
3. Make sure that the instrument is calibrated (refer
4. Follow the numbered steps in Fig. 7 to set the
scale limits.
Component Return Loss Limits
Antenna at Resonance
–14 dB
Connectors
–25 dB
Jumpers
–35 dB
Lightning Protector
–25 dB
Transmission Lines
–30 dB