Fault Location Mode
79
Figure 33 Measurement Hold
Markers
You must be in either Measure Match mode or Fault
Location mode to use markers.
You can activate and use up to six markers to visually
indicate the trace value at specific frequency points.
All markers follow the current trace only. The site
analyzer displays the frequency and signal value for
each marker at the bottom of the screen. You can
display each marker as a triangle pointer or a vertical
line through the trace. You can also display the
difference (delta) in frequency between two markers.
0.0
Meters
18.2
Calibration: Full
Save/
Recall
Auto
Scale
Limit
Test OFF
RUN
10.00
1.00
V
S
W
R
(
0.00/ )
Fault Location
[
]
238 Pts
Measurement Hold
Sweep will
NOT update
1
2
Step
Action
1
Press the Hold soft key to hold the mea-
surement; press the Run soft key to
resume sweeping
2
The lock symbol indicates that the mea-
surement is held