Bird Site Analyzer
20
Figure 3 Selecting the Measure Match Mode
Setting the Measurement Frequency
You must be in the Measure Match mode to set a
frequency.
Frequencies can be set manually or chosen from a list
of presets (band list). If the start, stop, center, or span
frequency is manually set, the band list will indicate
“Custom”. Be sure to set the frequency to a value that
is well beyond the normal range of the antenna. For
example, for a 450 MHz antenna, set the start
frequency at 400 and the stop frequency at 500 MHz,
for an 800 MHz antenna, set the start frequency at
700 and the stop frequency at 1,100 MHz.
1
2
3
800.00
MHz
1600.00
Calibration: OFF
Save/
Recall
Auto
Scale
Limit
Test OFF
HOLD
10.00
1.00
V
S
W
R
(0.00/
)
Measure Match
[238 Pts]
Step
Action
1
Turn On the Site Analyzer. Press ON/
OFF key for ½ second.
2
Press the MODE key
3
Press the Measure Match key
Summary of Contents for SITE ANALYZER SA-1700EX
Page 2: ...This page is not blank...
Page 8: ...Bird Site Analyzer vi...
Page 9: ...vii...
Page 30: ...Bird Site Analyzer 16...
Page 142: ...Bird Site Analyzer 128...
Page 148: ...Bird Site Analyzer 134...
Page 159: ...Maintenance 145 Measured Return Loss dB...
Page 167: ...Maintenance 153 the number of data points you wish measure 238 475 949...