INSTALLATION
Unpacking
•
The unit weighs ~13 kg. No special handling is required.
•
Please keep the original packaging for transport for service or repair as it has been specifically designed to protect the unit from damage
during transit.
•
Inspect the instrument and its power supply for any signs of damage caused during transit. If any damage is discovered, do not use the
instrument, and report the problem to your supplier.
Positioning
•
Ensure your proposed installation site conforms to the environmental conditions for safe operation:
•
Indoor use
•
5 to 40°C
•
Maximum relative humidity 90% up to 31°C decreasing linearly to 50% at 40°C.
•
Extremes of temperature may require recalibration of the unit for optimal performance.
•
The instrument must be placed on a stable, level bench or table capable of supporting its weight allowing sufficient space
around the instrument for air to circulate freely.
•
The instrument should be positioned so that the power supply cable may be readily removed in the event of a hazard or
malfunction.
•
Locate the instrument in an atmosphere free from dust and corrosive fumes. Use the dust cover to further protect the
instrument when not in use.
Installing
•
If the instrument has been stored in a cold environment, then it should be allowed to come to room temperature before
turning it on to avoid compromising the internal calibration procedure.
•
The equipment is operated using a 19 VDC power supply adapter unit. Always use the power supply adapter and mains cords
supplied with the instrument.
•
Mains power requirements are as follows:
•
100 to 240 VAC~
•
50 or 60 Hz
•
The UK style mains cord plug has a user replaceable 3A fuse. Replace only with the same rating and type 3A BS1362.
•
The unit maximum power rating is 90 VA.
•
Connect the instrument to the mains power using the main power cord and the 19 VDC power supply adapter unit, then turn
the instrument’s
main switch to the on (I) position, this will Power on the instrument followed by a series of self-diagnostic
checks.
Main switch and 19 VDC power supply socket
Summary of Contents for HB Ultrospec 7500
Page 1: ...ORIGINAL INSTRUCTIONS 41 56 2050 REV03 1 Ultrospec 7500 Spectrophotometer USER MANUAL...
Page 2: ...Ultrospec 7500 User Manual 41 56 2050 REV03 2 Intentionally blank...
Page 6: ...Ultrospec 7500 User Manual 41 56 2050 REV03 6 Intentionally blank...
Page 103: ...Ultrospec 7500 User Manual 41 56 2050 REV03 103 Intentionally blank...
Page 104: ......