Ultrospec 7500 User Manual
•
41 56 2050 REV03
•
35
Step 9
Proceed to the next parameters screen using the right/forward arrow.
Step 10
Set the mode to define the desired result
, “Delta A”, “Final A”, or “Slope”.
Step 11
Set the factor value to be applied to the result, between 0.000 and ±9999.
Step 12
Define the units that the result value will be reported in.
Step 13
Set the y-axis minimum and maximum to between -4 and 4.
Step 14
Proceed to the next parameter screen using the right/forward arrow.
Step 15
Set auto print to “On” or “Off”. If auto print is set to “On”, select the print
to hardware from “Internal Printer”, “PC via USB”,
or
“USB Mass Storage”
depending on what hardware is connected to the instrument.
Step 16
Set auto save to “On” or “Off”.
If auto save is set to “On”, select the save
to hardware from “USB CSV”, “USB”, or “Internal” depending on what
hardware is connected to the instrument.
Step 17
Proceed to the next parameter screen using the right/forward arrow.
Step 18
Set whether to use as single cell to “On” or “Off”. If set to “Off”, set the
position prompt per sample to “On” or “Off”.
Step 18
Proceed to the measurement screen using the right/forward arrow.
If using a single cell holder, or a cell changer set to use as a single cell
holder, skip step 19 and go straight to step 20.
Summary of Contents for HB Ultrospec 7500
Page 1: ...ORIGINAL INSTRUCTIONS 41 56 2050 REV03 1 Ultrospec 7500 Spectrophotometer USER MANUAL...
Page 2: ...Ultrospec 7500 User Manual 41 56 2050 REV03 2 Intentionally blank...
Page 6: ...Ultrospec 7500 User Manual 41 56 2050 REV03 6 Intentionally blank...
Page 103: ...Ultrospec 7500 User Manual 41 56 2050 REV03 103 Intentionally blank...
Page 104: ......