Ultrospec 7500 User Manual
•
41 56 2050 REV03
•
7
ESSENTIAL SAFETY NOTES
Hazards and Warnings
This section describes potential hazards which may exist in the operation of these units. Several warning labels and symbols are affixed to
your instrument. These symbols are used to inform you of potential dangers which may exist or where caution is required. Before installing
your new unit, please take time to familiarise yourself with these warnings and symbols.
N.B. THE PROTECTION GIVEN BY THE EQUIPMENT MAY BE IMPAIRED IF USED IN A MANNER NOT SPECIFIED BY THE MANUFACTURER.
This instrument is subject to the following identified hazards:
This unit uses a Xenon lamp. The lamp energy is mainly confined within the unit but traverses the cell holder when a
measurement is being taken. Although the energy present is low and intermittent you are advised not to stare into the beam
or attempt to deflect the beam as prolonged exposure could result in permanent eye damage.
High voltages exist within the power supply unit and the Xenon lamp housing. Repair and maintenance should only be carried
out by individuals trained to work on these instruments.
There are no biohazardous materials within the unit, however, this unit may be exposed to biohazardous samples during
normal laboratory use. To protect users against these hazards we recommend the following decontamination procedures:
•
Wipe the exterior casework with disinfectant cleaning wipes.
•
Remove cuvettes and cuvette holders.
•
Wash with disinfectant appropriate for the biohazard in question.
•
Rinse with distilled water.
•
Allow to dry thoroughly before reuse.
To further reduce the possibility of biohazards:
•
Include an appropriate decontamination certificate for equipment returned for repair.
•
Ensure that the operator of the equipment is provided with a safe working environment.
•
Use, store and dispose of any chemicals in accordance with manufacturer’s guidelines and local safety regu
lations.
•
Provide suitable ventilation when working with volatile solvents or toxic substances.
•
Dispose of solvents and chemicals that may be classed as hazardous waste in accordance with local regulatory practice.
•
Determine if personal protective equipment (PPE) is required for handling laboratory samples.
All models can be connected to and operated from a PC. To preserve the integrity of the measuring equipment it is essential
that the attached PC itself conforms to basic safety and EMC
standards and is set up in accordance with the manufacturers’
instructions. If in doubt, consult the information that came with your PC.
The following safety precautions should be observed when operating a PC:
•
To reduce the chance of eye strain, set up the PC display with the correct viewing position, free from glare and with
appropriate brightness and contrast settings.
•
To reduce the chance of cross contamination from biological samples, use appropriate personnel protection measures
and disinfectant wipes on keyboard and mouse.
In the event of contamination, malfunction or hazard occurring, the operator should disconnect the unit, by removing the
power cord, and isolate for decontamination and/or repair.
Summary of Contents for HB Ultrospec 7500
Page 1: ...ORIGINAL INSTRUCTIONS 41 56 2050 REV03 1 Ultrospec 7500 Spectrophotometer USER MANUAL...
Page 2: ...Ultrospec 7500 User Manual 41 56 2050 REV03 2 Intentionally blank...
Page 6: ...Ultrospec 7500 User Manual 41 56 2050 REV03 6 Intentionally blank...
Page 103: ...Ultrospec 7500 User Manual 41 56 2050 REV03 103 Intentionally blank...
Page 104: ......