Equation Editor
The Equation Editor application is selected from the Applications screen. It can be used to create more complex custom methods
incorporating bespoke calculations.
Step 1
Select the mode from “Absorbance” or “%Transmission”.
Step 2
Set sample
message prompt between λ to “On” or “Off”
.
Step 3
Set scan to “On” or “Off”.
Step 4
Proceed to the next parameter screen using the right/forward arrow.
Step 5
To define the measurements to be taken per sample:
•
Under the “Name” column, allocate
a name to the measurement.
•
Under the “λ”
column, set the wavelength to between 190 and 1100 nm,
that the measurement will taken at.
•
Under the “Function” column, select the data
extracted by
measurement from “Abs/%T at λ”, “Peak closest to λ”, or “Valley closest
to λ”.
•
For the peak and valley functions, under the “+/
-
“ column select the
search tolerance from “1nm”, “2nm”, “5nm”, “10nm”, or “20nm”.
Step 6
Proceed to the next parameter screen using the right/forward arrow.
Step 7
Select the sample Id mode
from “
Default
”, “Auto Increment”, “
Prompt for ID
”
,
or “Fixed List”
.
Step 8
If the sample Id is set to “Default”, “Auto Increment”, or “Prompt for ID” skip
steps 8 through 10 and go to step 11.
If the sample Id is set to “Fixed List”, define the number of samples to between
1 and 99.
Step 9
Proceed to the next parameter screen using the right/forward arrow.
Step 10
E
nter the sample Ids in the “Sample Names” table. If the list comprises
multiple pages, us the up and down page arrows to change the page:
Summary of Contents for HB Ultrospec 7500
Page 1: ...ORIGINAL INSTRUCTIONS 41 56 2050 REV03 1 Ultrospec 7500 Spectrophotometer USER MANUAL...
Page 2: ...Ultrospec 7500 User Manual 41 56 2050 REV03 2 Intentionally blank...
Page 6: ...Ultrospec 7500 User Manual 41 56 2050 REV03 6 Intentionally blank...
Page 103: ...Ultrospec 7500 User Manual 41 56 2050 REV03 103 Intentionally blank...
Page 104: ......