Step 19
Press the batch measurement icon, then load the cell changer according
to the cell changer prompt. Then select the confirm icon.
If using a cell changer, skip steps 20 through 21 and go straight to step 22.
Step 20
Insert the reference sample then take a reference measurement using the
reference measurement icon.
The acquired reference sample baseline will be applied to all subsequent
sample measurements.
Step 21
Replace the previous sample with a test sample then take a sample
measurement using the sample measurement icon.
Repeat for all samples.
Step 22
If sample overlays are active, the sample overlay tools will appear to the right of the displayed spectra.
Select data
Select the sample overlay to be analysed
Display data
Hide or show sample overlays
Delete data
Delete sample overlays from the overlay tools list
Additional viewing tools as available at the bottom of the screen.
Cursor left
Move the x-axis cursor position left
Cursor right
Move the x-axis cursor position right
Zoom in
Zoon in to the area around the x and y-axis cursor position