Appendix B Product Specifications
--158--
User Manual of Veterinary Monitor
Unit selection
%, mmHg, kPa
Operating temperature
0
℃~
40
℃
(32
℉~
104
℉
)
Storage&Transport temperature
-40
℃~
75
℃
(-40
℉~
167
℉
)
Operating humidity
10 %
~
95 % (non-condensing)
Storage&Transport humidity
5 %
~
100 % (non-condensing)
Operating atmospheric pressure
52.5 kPa
~
120 kPa (393.75 mmHg
~
900 mmHg)
Storage&Transport atmospheric pressure
50 kPa
~
120 kPa (375 mmHg
~
900 mmHg)
Total System Response Time
<
1 s
Primary agent threshold
0.15 vol%. When an agent is identified, concentrations will be reported even below 0.15
vol% as long as apnea is not detected.
Secondary agent threshold
0.2 vol% + 10% of total agent concentration
Range of Breathing Rate
0 rpm
~
150 rpm
Accuracy of Breathing Rate
±1 rpm
Asphyxia Alarm Delay
20 s, 25 s, 30 s, 35 s, 40 s, 45 s, 50 s, 55 s, 60 s
Automatic Pressure Compensation
yes
Alarm Limit Specifications
Range
EtCO
2
Upper Limit
(Lower Limit +2)mmHg
~
99 mmHg
EtCO
2
Lower Limit
0 mmHg
~
(Upper Limit -2)mmHg
FiCO
2
Upper Limit
0 mmHg
~
99 mmHg
awRR Upper Limit
(Lower limit+2) rpm
~
100 rpm
awRR Lower Limit
0 rpm
~
(Upper limit-2) rpm
Respironics Sidestream LoFlo Module
Measuring Method
Infrared Spectrum Method
Measuring Mode
Sidestream
Preheating time
Max. length of waveform is 20s. Full accuracy requirements satisfied after 2min
(environment temp.: 25
℃
)
Range
0%
~
19.7% (0 mmHg
~
150 mmHg) (0 kPa
~
20 kPa)
Resolution
0.1 mmHg 0 mmHg
~
69 mmHg
0.25 mmHg 70 mmHg
~
150 mmHg
Stability
Short-term drift:
≤
0.8 mmHg (0.1 kPa) within 4h
Long-term drift: accuracy maintained within 120h.
Unit selection
%, mmHg, kPa
Operating temperature
0
℃~
40
℃
(32
℉~
104
℉
)
Storage temperature
-40
℃~
70
℃
(-40
℉~
158
℉
)
Operating humidity
10 %
~
90 % (non-condensing)
Storage humidity
10 %
~
90 % (non-condensing)
Storage atmospheric pressure
53.33 kPa
~
106.67 kPa (400 mmHg
~
800 mmHg)
Accuracy (Gas Temp. at 25
℃
)
0 mmHg
~
40 mmHg (0 kPa
~
5.3 kPa), ±2 mmHg (0.27 kPa)
41 mmHg
~
70 mmHg (5.5 kPa
~
9.3 kPa), ±5% of the reading
71 mmHg
~
100 mmHg (9.4 kPa
~
13.3 kPa), ±8% of the reading
101 mmHg
~
150 mmHg (13.4 kPa
~
20 kPa), ±10% of the reading
(When the breathing rate is
>
80 rpm, all ranges are ±12% of the reading)
Total System Response Time
<
3 s
Range of Breathing Rate
2 rpm
~
150 rpm
Summary of Contents for iM 12
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Page 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Page 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Page 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Page 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Page 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Page 195: ......