Chapter 6 Parameter Monitoring
--72--
User Manual of Veterinary Monitor
Warning
Please regularly check the color, warmth and sensitivity of the animal
‟s distal limbs in the process of
measuring blood pressure under Auto and STAT modes. Once any abnormality is found, please
relocate the cuff or terminate blood pressure measure.
Interval Setup
Select [Interval] in the menu [NIBP Setup] when the measure mode is Auto, under the auto mode, the
monitor automatically and repeatedly conducts NIBP measures according to the preset time intervals.
Attention
This option is available only when the [Measure Mode] is Auto.
Alarm Setup
Select [Alarm Setup>>] in the menu [NIBP Setup] and then set in the pop-up menu the alarm attributes of
each NIBP option parameter.
Start /Stop NIBP Measure
Select this option to start or stop NIBP measure in accordance with the preset mode.
Stop All.
Select this option at any time when the measure is in process to terminate all measuring tasks.
Venipuncture
Users can inflate with NIBP cuff to produce a pressure appropriate for blocking venous vessels and thus
assisting venipuncture.
1.
Select [Venipuncture>>] in the window [NIBP Setup] and then set [Cuff Pressure] in the pop-up menu
for values for venipuncture pressure.
2.
Select [Venipuncture]
3.
Puncture venous vessel and take blood samples.
4.
Select the NIBP button in the front panel or [Stop All.] in the window [NIBP Setup] to deflate the cuff;
if it is not deflated, the cuff will be automatically deflated when a preset period has elapsed.
The cuff inflation pressure and the remaining venipuncture time will be displayed on the NIBP parameter
region during puncture.
NIBP List Review
Select this option to enter the [NIBP List] window directly. Please refer to 8.4.5 for more details.
6.5.8 NIBP Leakage Test
The system gives no prompt when the leakage test is passed, while a corresponding message will be displayed in
the non-invasive blood pressure (NIBP) parameter region if failed:
Leakage Test Procedure:
1
Connect the cuff and the monitor‘s blood pressure port.
2
Wrap the cuff onto a column of proper size, as referred to Figure 6.32
3
Select [Main Menu]-[Maintenance]-[NIBP Leakage Test], and then ‗Leakage Testing
…
‘ is displayed in the
NIBP parameter region, indicating the system has started conducting leakage detection.
4
The system automatically inflates until the pressure is 180 mmHg (24 kPa).
5
The system automatically open deflation valve about in 20 seconds, indicating leakage detection is
completed.
Summary of Contents for iM 12
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Page 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Page 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Page 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Page 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Page 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Page 195: ......