Chapter 6 Parameter Monitoring
--82--
User Manual of Veterinary Monitor
module
‟s service life will occur.
Excessive negative or positive pressure in the animal
‟s tube may affect sampling flow rate.
Please insert the sampling tube before connecting a T-valve in the respiratory loop. Please remove
the T-valve from the respiratory loop before removing the sampling tube.
Brief Description on Measure
This series of monitors allow selection of different side-stream CO
2
modules, and the measuring method for
side-stream CO
2
module will be discussed in the following by taking the Kingst Side-Stream KM7002-V33
module as the example.
1.
Attach a water tank to a base firmly and connect the CO
2
measuring assembly, shown in Figure 6.44
Figure 6.44 Dewatering Bottle Attaching and Detaching Diagram
2.
As shown in Figure 6.44, hold the two ends of dewatering bottle and gently pull the tube down to
remove water.
3.
As shown in Figure 6.45, connect one end of the sampling tube with the dewatering bottle‘s threaded
nipple and the other end with the breathing tube of the animal‘s breathing machine or anesthesia
machine or with the threaded nipple having a diameter of φ10mm in other airways (a section of
connecting tube having a φ10mm threaded nipple needs to be connected if such source nipple is not
available), or directly fixing the sampling tube to the animal‘s nostrils by using adhesive tape.
Figure 6.45 Dewatering Bottle Connecting Diagram
Attention
Pay close attention to the water level in the dewatering bottle and make sure it is below the highest
level; and exchange dewatering bottles to prevent water entering the module.
Summary of Contents for iM 12
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Page 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Page 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Page 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Page 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Page 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Page 195: ......