Chapter 11 Other Functions
--132--
User Manual of Veterinary Monitor
Figure 11.2 Nurse Call Setup
Nurse Call
Set nurse call function: Off or On.
Signal Type
Set the type of output signal that triggers nurse calling system: Pulse Low or Hold Low.
Trigger Alarm Level
Set the alarm level that triggers nurse calling message: High, Mid, Low, you can choose multiple or none.
Trigger Type
Set the alarm type that triggers nurse calling message: Physiology alarm or Technology alarm, you can
choose multiple or none.
Attention
If you select neither [Trigger Alarm Level] or [Trigger Type], then no alarm will trigger nurse calling
signal.
Warning
Nurse call function cannot serve as the main alarm message source. You should combine the
audible and visual alarm of the equipment with animals
‟ clinical manifestations and symptoms, to
serve as the main alarm message source that notifies the doctors and nurses about an alarm state.
11.5 Defibrillation synchronization
When detecting R wave, the monitor will output a defib.Syn pulse signal. Defibrillator gets the defib.Syn signal
from the monitor and should discharge accordingly during the declining period of R wave to reduce damage to the
animal‘s heart.
Warning
Improper use of a defibrillator may cause injury to the animal. The user should determine whether to
perform defibrillation or not according to the animal
‟s condition.
Before defibrillation, the user must ensure both defibrillator and monitor has passed the system test
and can be safely used jointly.
Before defibrillation, make sure that [Defib. Sync] is set to [On] and the [Filter] is set to [Diagnostic].
After defibrillation is finished, disconnect the defibrillator.
Summary of Contents for iM 12
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Page 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Page 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Page 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Page 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Page 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Page 195: ......