Chapter 6 Parameter Monitoring
--104--
User Manual of Veterinary Monitor
device>, causing the ISA/<host device> to fall on the animal.
Dispose nomoline family sampling lines in accordance with local regulations for biohazardous
waste.
Use only airway T-adapters with the sampling point in the center of the adapter.
Do only use sample lines intended for anesthetic agents if N
2
O and/or anesthetic agents are being
used.
Do not use T-adapter with animal < 10 kg, as this adds 7 ml dead space to the animal circuit.
Do not use the ISA gas analyzer with metered-dose inhalers or nebulized medications as this may
clog the bacteria filter.
Since a successful zeroing requires the presence of ambient air (21% O
2
and 0% CO
2
), ensure that
the ISA is placed in a well ventilated place. Avoid breathing near the ISA sidestream gas analyzer
before or during the zeroing procedure.
Never sterilize or immerse the ISA sidestream gas analyzer in liquid.
Measurements can be affected by mobile and portable RF communications equipment. Make sure
that the ISA sidestream gas analyzer is used in the electromagnetic environment specified in this
manual.
ISA sidestream gas analyzer is intended only as an adjunct in animal assessment. It must be used
in conjunction with other assessments of clinical signs and symptoms.
Replace the sampling line if the sampling line input connector starts flashing red, or a
“Sample line
clogged” message is displayed on the host.
No modification of this equipment is allowed without authorization of the manufacturer. If this
equipment is modified, appropriate inspection and testing must be conducted to ensure continued
safe operation.
ISA sidestream gas analyzers are not designed for MRI environments.
During MRI scanning, the <host device> must be placed outside the MRI suite.
Use of high frequency electrosurgical equipment in the vicinity of the ISA/<host device> may
produce interference and cause incorrect measurements.
Do not apply negative pressure to remove condensed water from the nomoline family sampling line.
Too strong positive or negative pressure in the animal circuit might affect the sample flow.
Strong scavenging suction pressure might affect the sample flow.
Exhaust gases should be returned to the animal circuit or to a scavenging system.
Due to the risk of animal cross-infection, always use a bacteria filter on the exhaust port side if
sampled gas is intended to be re-breathed.
Do not place the ISA gas analyzer in any position that might cause it to fall on the animal.
Do not re-use disposable single-animal use nomoline family sampling lines due to the risk of cross
contamination.
Do not sterilize or immerse nomoline family sampling lines in liquid.
Do not operate the ISA sidestream gas analyzer if the enclosure is damaged.
Attention
The ISA
“plug-in and measure” analyzers should be securely mounted in order to avoid the risk of
damage to the ISA.
Do not operate the ISA sidestream gas analyzer outside the specified operating environment.
(US Only) Caution: federal law restricts this device to sale by or on the order of a licensed
healthcare practitioner.
Summary of Contents for iM 12
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Page 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Page 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Page 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Page 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Page 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Page 195: ......