ONCORE PRO, Operating Manual, Rev. K
15
Section 3.
Installation Requirements and Instructions
This section provides an overview of the installation of the Oncore Pro Slide Stainer to be performed by the field
service engineer. The following requirements should be met to ensure proper instrument function.
3.1.
Installation Requirements
Location
Place the system in an area where it is easy to operate and easy to connect to the instrument.
Surface
Prior to unpacking, ensure that the work area for the instrument is a solid, level surface that can safely
support the weight of the system.
Size
Minimum dimensions for the work area are 36.5 inch (93 cm) width x 36 inch (91 cm) height x 24 inch (61 cm)
depth clearance.
Environment
The work area must be in an environment with an ambient temperature between 18°C-26°C (64°F-
79°F). The area must be properly ventilated with a maximum relative humidity of 80% for temperatures up to
31
o
C; Pollution Degree II or less. The equipment will produce less than 85 decibels at peak operation.
Power source
The instrument and computer must be connected to a 120V/240 15A supply with grounding (3
prong). It is recommended to use a dedicated power source to prevent interference from other instruments or
equipment. It is also recommended to use a UPS (uninterruptible power supply) with minimum 1500VA. This will
help to prevent unexpected system failures due to an unstable power grid system or in the event of a temporary
power outage.
3.2.
Installation Instructions
Unpacking
The system is packaged in a wooden crate mounted on a pallet. Unlatch all fasteners and remove the
lid on top of the crate first. Remove the computer, accessories box and label printer from the shelf inside the
crate. Remove the shelf and the sides of the crate.
Installation
Requires 2 people. Each person should pick up one end of the instrument and carefully lift it out of
the crate. Carefully move the instrument to the operating area and place it on the bench. Level the instrument
surface using a bubble level. Remove all packing material and the packaging securing the robotic arm. Verify that
all axes can move freely before operating the instrument. Unpack and install the computer and accessories.
Hooking up
Connect the USB cable from the designated port on the computer to the back of the instrument.
Do not plug and unplug the USB cable while the PC or Instrument is on.
***Always turn off the instrument and PC before plugging and unplugging the USB cable.
Connect the Computer to the power source or UPS. Follow the instructions on the UPS, as some may require
powering up for several hours to charge the internal battery. Using a standard IEC320 compatible power cord,
connect the AC power port at the rear of the instrument to the power source or UPS.
Cut the waste tubing to size and route to the waste bottles on the floor. The ideal configuration is straight down to
the waste bottle in order to minimize the chance of overflow and waste buildup.
Avoid immersing the waste tubing in the waste bottle or flattening and kinking the tubing as this
may disrupt the gravity fed drainage and result in waste overflow.