To read the information in the RFID tag
, click “
Read Tag
”. The programmed information will be displayed in the
empty fields to the left.
Note: Brand new RFID tags will not be read correctly, as they only contain a randomly generated Tag ID.
To write information into the RFID tag
, complete the entry fields to the right and click “
Write Tag
”.
Select the
Reagent Type
and
Reagent Name
from the dropdown lists.
Note: The user may only select user-added reagents.
Some reagents may not appear on the user-programmable reagent list.
Select the appropriate
Bottle Type
(15 or 7 mL).
The bottle type is required to ensure accurate liquid detection and number of tests.
Enter the
Storage Temperature
.
Enter the
Number of Tests Left
.
15 ml vials should be tagged as 60 tests. 7 ml vials should be tagged as 30 tests.
Vials must be filled properly to the line indicated.
Note1: The system may permit the user to reuse/reprogram user-programmed vials.
Please ensure vials are thoroughly cleaned and dried before re-use.
Note2: The system may block access for reprogramming reagent kit vials.
Enter the
Lot Number (up to 12 characters)
and
Expiration Code (yyyymm)
.
(Optional) Select the
Manufacturer Name
.
Note: Select “---Add or Remove---“, from the dropdown menu to add a new reagent manufacturer name.
Enter the
Catalog Number
.
The system will immediately read the tag after writing to verify and display the data in the left field.
Do not remove the vial until the tag ID on the left is highlighted in green to indicate that the tag
has been read successfully.