15.5 Self-test function (MK 56)
The self-test function
enables an automated check of the proper chamber functioning as well as a target-
ed and reliable fault analysis
. It is available with the “Master”, “Service”, and “Admin” authorization levels.
In this case, the chamber successively undergoes various defined operating states, which serves to de-
termine reproducible characteristic values. These characteristic values provide information on the perfor-
mance and precision of the individual functional systems of the chamber (e.g., heating, refrigeration) of
the chamber.
The results of the self-test
are stored in the service recorder of the controller
. You can export them using
the controller’s USB interface and send them to BINDER Service (use function “Export service data” to
USB stick, chap. 14.6). BINDER Service will evaluate the data using an analyzing tool.
Activating the self-test mode
In order to allow an optimum comparison of the determined characteristic values with the
reference characteristic values, the ambient temperature should be in the range of +22 °C
+/- 3 °C /
71.6 °F +/- 5.4 °F
.
The chamber shall be unloaded (empty with standard equipment).
Path:
Main menu > Settings > Various
Submenu “Various”.
Scroll all the way down
to access the “Self-test”
function.
Submenu “Various”.
Select the field “Self-test”.
Submenu “Various”.
To start the self-test, select the desired test duration.
Confirm entry with
Confirm
icon.
Return to Normal display with the
Back
icon to take
over the entries.
Summary of Contents for MK 115
Page 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Page 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Page 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Page 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Page 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Page 144: ...MK MKT E5 03 2019 page 144 158 ...
Page 145: ...MK MKT E5 03 2019 page 145 158 ...
Page 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Page 147: ...MK MKT E5 03 2019 page 147 158 ...
Page 148: ...MK MKT E5 03 2019 page 148 158 ...
Page 150: ...MK MKT E5 03 2019 page 150 158 ...
Page 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Page 159: ......