MK / MKT (E5) 03/2019
page 103/158
17. Notes on refrigerating operation
Defrosting:
BINDER chambers are very diffusion-proof. To ensure high temperature precision there is no automatic
cyclic defrosting device. The refrigerating system largely avoids icing of the evaporation plates. However,
at very low temperatures the moisture in the air can condense on the evaporator leading to icing.
Always close the door properly.
Operation with temperature set-points above +5 °C / 41 °F at an ambient temperature of 20 °C /
68 °F:
The air defrosts the ice cover automatically. Defrosting is continually performed.
Operation with temperature set-points below +5 °C / 41 °F:
Icing on the evaporator is possible. Defrost the chamber manually.
With temperature set-points below +5 °C /
41°F,
regularly defrost the chamber manually:
•
Set the temperature to 60 °C /
140°F
(Manual Mode).
•
Let the chamber operate for approx. 1 hour with the door closed. Remove the access port
plugs.
Too much ice on the evaporator is noticeable by reduced refrigerating performance.
Operation with temperature set-points below 0 °C / 32 °F:
While operating the chamber with set-points below 0 °C /
32 °F
condensation is possible at the inner sur-
face of the door around the door gasket.
In case of heavy condensation, check tightness of the door gasket.
After one or two days operation at a set-point below 0 °C /
32 °F
a thin ice layer can cover the inner cham-
ber door, the front margins of the inner kettles and may be the glass window.. The amount depends on the
ambient temperature and humidity. This does not influence the proper function of the refrigerating system.
Refrigerating performance decreases while operating the chamber at temperatures below 0 °C
/
32 °F
due to icing of the evaporators. For this reason defrost the chamber regularly, e.g. once
a week.
CAUTION
Uncontrolled defrosting of icing on the evaporator.
After several days of refrigerating below +5 °C /
41 °F
:
∅
Do NOT directly turn off the chamber.
Manually defrost the chamber (see description above).
Then, shut down the chamber at the main power switch (1). Keep removed the access
port plugs.
Summary of Contents for MK 115
Page 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Page 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Page 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Page 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Page 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Page 144: ...MK MKT E5 03 2019 page 144 158 ...
Page 145: ...MK MKT E5 03 2019 page 145 158 ...
Page 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Page 147: ...MK MKT E5 03 2019 page 147 158 ...
Page 148: ...MK MKT E5 03 2019 page 148 158 ...
Page 150: ...MK MKT E5 03 2019 page 150 158 ...
Page 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Page 159: ......