9.7.2 Set-point ramp and set-point step
You can define the type of temperature transitions for each individual program section.
“Ramp” mode: Gradual changes of temperature
The set-point of a given program section functions as the section’s start temperature. During the section’s
duration, the set-point gradually passes to the set-point of the subsequent program section. The actual
value follows the continually changing set-point.
If the last program section is in “ramp” mode and the setpoint shall change within this section, then you
must program an additional section (with the shortest possible section duration) to provide the target tem-
perature of the last program section. Otherwise, the setpoint would remain constant during the section’s
duration.
Programming in the “ramp” mode allows all kinds of temperature transitions:
•
Gradual changes of temperature
The setpoint changes its value gradually during the entered section duration. The actual value follows
the continually moving set-point at any time.
•
Program sections with constant temperature
The setpoints (initial values) of two subsequent program sections are identical; so the temperature re-
main constant during the entire duration of the first program section.
•
Sudden changes of temperature
Steps can be programmed in ramp mode as temperature changes (ramps) that occur during a very
short interval. If the duration of this transitional program section is very short (minimum entry 1 sec),
the temperature change will proceed rapidly within the minimum amount of time.
“Step” mode: Sudden changes of temperature
The set-point of any program section functions as the section’s target value. At the start of the program
section, the unit heats up or cools down the chamber with the maximum speed to reach the entered value;
and then it holds it for the remaining section time. Therefore the set-point temperature remains constant
for the section’s duration. These changes occur rapidly within the minimum amount of time (minimum
entry: 1 second).
Programming in the “step” mode allows only two kinds of temperature transitions:
•
Programming gradual changes of temperature (ramps) is impossible in the “step” mode
•
Program sections with constant temperature
The setpoints (target values) of two subsequent program sections are identical; so the temperature
remain constant during the entire duration of the first program section.
•
Sudden changes of temperature
The entered setpoint of the section is reached as fast as possible and then held constant for the re-
maining section duration.
Selecting the setting “Ramp” or “Step”
Section view (partial view).
In the field “Course” select the desired setting “Ramp”
or “Step”.
Summary of Contents for MK 115
Page 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Page 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Page 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Page 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Page 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Page 144: ...MK MKT E5 03 2019 page 144 158 ...
Page 145: ...MK MKT E5 03 2019 page 145 158 ...
Page 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Page 147: ...MK MKT E5 03 2019 page 147 158 ...
Page 148: ...MK MKT E5 03 2019 page 148 158 ...
Page 150: ...MK MKT E5 03 2019 page 150 158 ...
Page 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Page 159: ......