MK / MKT (E5) 03/2019
page 27/158
4.4.2 Connecting the voltage and frequency changer (for chambers equipped with a
voltage and frequency changer)
The voltage and frequency changer is supplied with a fixed power connection cable without a plug. It is
protected against excess-current with 3 internal overload releases. The connection is made by the cus-
tomer.
The socket must provide a protective conductor.
Electrical connection data:
•
Input side: 480 V, 60 Hz, 4-wire
•
Output side (to the chamber): 400 V, 50 Hz, 5-wire
To establish the electrical connection of the alternating climate chamber with the voltage and frequency
changer, proceed in the following order:
1. Connect the chamber to the connection socket (G) of the voltage and frequency changer
2. Establish the power connection of the voltage and frequency changer using the power cable (I)
3. Turn on the voltage and frequency changer at the power switch (H) (position “ON”)
4. Turn on the chamber with the main power switch (3) in the lateral control panel
Left side of the voltage and frequency changer
with connection socket (G) for the alternating climate
chamber
Right side of the voltage and frequency changer
with power switch (H) and power cable (I)
Figure 12: Voltage and frequency changer, mounted
(H)
(I)
(G)
Summary of Contents for MK 115
Page 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Page 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Page 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Page 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Page 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Page 144: ...MK MKT E5 03 2019 page 144 158 ...
Page 145: ...MK MKT E5 03 2019 page 145 158 ...
Page 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Page 147: ...MK MKT E5 03 2019 page 147 158 ...
Page 148: ...MK MKT E5 03 2019 page 148 158 ...
Page 150: ...MK MKT E5 03 2019 page 150 158 ...
Page 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Page 159: ......