Patient Thermoregulation – Step by Step Operation
Preparing the System for Operation
To prepare the system for operation:
1. Place the unit in the desired position according to “Space and
Environmental Requirements”.
2. Remove the water tank feeder cover and pour in only sterile water
until the maximum allowable level is reached (minimum water
temperature 13ºC / 55.4ºF).
3. Observe the water-level indicator to prevent overfilling the water tank.
Close the water tank feeder cover.
4. Connect CritiCool
®
MINI to the power source.
NOTE:
Use only sterile water or 0.22 micron filtered tap water.
NOTE:
In case of overfilling, see Table 6: Technical Messages and Alarms.