7. Wait for the water to drain from the system. When the water has
been emptied completely, a message appears stating that the
CritiCool
®
MINI is now ready for storage until the next procedure.
Replacing the Wrap
WARNING!!!
Avoid disconnecting tubes above electrical equipment
as mild dripping may occur during disconnection.
To replace the Wrap:
1. Switch to STANDBY and wait a minute to let the water return to
the system.
2. Close wrap clamps to avoid water spill.
3. Disconnect the connecting tubes from the wrap.
4. Remove the used wrap and dispose of according to hospital
regulations.
5. Position the new wrap (follow the instructions for use leaflet
supplied with each wrap).
6. Reconnect the connection tubes to the new wrap.
7. When wrapping the patient, follow the instructions for use in the
leaflet included with each wrap.
Figure 29: Emptying Water-Performing Panel