Set-Point / Target Temperature Set Up
The set-point is the chosen temperature in TTM and Normothermia to
which the thermoregulation system cools or warms the patient's body.
The Target Temperature is the chosen temperature in controlled
rewarming to which the thermoregulation system warms the body to return
the patient to a normothermic temperature.
NOTE:
Upon start up, the default set-point for
TTM Mode
is 33.5°C
(92.3°F).
Upon start up, the default set point for
Normothermia Mode
is
36.5°C (97.7°F).
After start up, it is possible to change both the Set Point and Target
Temperature.
To change the Set Point /Target Temperature
1. Touch the Set Point/Target Temperature icon to display the Set
Point/Target Temperature setting screen panel.
Figure 25: Set-Point Setting Screen
2. Use
and
to select the Set Point/Target Temperature to provide a
change of 0.1°C. Each tick mark in the scale provides a change of 1.0°C.
3.
When finished, touch
OK
.