CAUTION!
•
Do not use any kind of brush on the machine touch screen or its
accessories.
•
Do not submerse the machine in liquid.
•
Do not wash the electrical power socket.
•
Do not use any saline or irrigated fluids.
•
Do not use any ester solvents.
For reusable temperature probes, follow the manufacturer’s recommendations
and always check the temperature probes for scratches and tears before and
after cleaning. If the probe is damaged, do NOT use it.
NOTE:
Follow your hospital protocols for disinfecting the product.
Required Tools for Cleaning and Disinfection
•
PPE (Personal Protective Equipment) according to the disinfectant
manufacturer's instructions.
•
Lint free cloths.
•
Sodium Dichloroisocyanurate (NaDCC) powder or tablets
•
Sterile water / 0.22 micron filtered tap water (approximately 1.2 liters)
Recommended Disinfectants for External Surfaces
•
Chlorinated bleach solution (5.25% sodium hypochlorite concentration)
•
Quaternary ammonium compounds (ammonium chloride as active
ingredient)
Before Each Use
CAUTION!
Apply finger pressure only. External instruments exert excessive
pressure on the screen and should not be used.
1. Use PPE as recommended by the disinfectant’s manufacture.
2. Make sure that the system is turned off and unplugged from power.
3. Using a lint free cloth with sterile water, clean the exterior of the
machine and the LCD screen from any soiling.
4. Prepare the disinfectant solution as described by the manufacturer and
follow the manufacturer’s directions for time duration and
concentration.
5. Using a lint free cloth with the disinfectant, disinfect the exterior of the
machine, the LCD screen, and the hoses.