DDT-320-000 Rev G Belmont Medical Technologies
Page
91
of
114
Figure 47: CritiCool MINI System Message Troubleshooting Guide
Message
Cause of Problem
Action to
be taken
Comments
Water Temperature Too High
When the water
temperature in the
system is more than
42°C (107.6°F).
Thermoregulation
stops until the
water cools or the
system halts.
Turn OFF the
system for 3
seconds and then
turn it ON again. If
the problem
persists,
turn OFF CritiCool
®
MINI and contact
your Belmont
Medical
Technologies
representative.
The alarm can be muted for
unlimited time.
Patient’s temperature is above
XX.X°C
The alarm for High
Patient Temperature
can be configured in
“settings”. The alarm
and message are
issued according to the
selected alarm limit.
The available values
are: 36°C, 36.5°C,
37°C, 37.5°C, 38°C,
38.5°C
Check that the
core temperature
probe is in place
and follow the
patient’s
temperature.
Inform the
clinician.
Thermoregulation
continues.
The alarm can be muted for
30 minutes.
Low Core Temperature
Thermoregulation is Continuing..
This message appears:
1. When Core temp. is
>0.8°C less than the
Set Point.
2. According to
alarm settings.
Check that the
core temperature
probe is in place
and keep following
the patient’s
temperature.
No action is
required.
If rewarming
manually: Do not
attempt to
increase more
than 0.8°C above
actual core
temperature.
For this message, an alarm
issues but thermoregulation
continues.
The alarm can be muted for
30 minutes.