DDT-320-000 Rev G Belmont Medical Technologies
Page
80
of
114
1. In the main menu, select
Services
.
Figure 37: Selecting Thermal Disinfection Service
2. Touch
Thermal Disinfection
and then
OK
.
3. The process is password protected. Enter password.
4. Touch OK. The following verification message appears:
5. Verify that the tank is full. Connect the bypass tube and touch OK.
Thermal Disinfection begins. Countdown appears on the screen.
The process takes about 2 to 3 hours.
Figure 38: Initiating Thermal Disinfection