DDT-320-000 Rev G Belmont Medical Technologies
Page
84
of
114
Filter Replacement
The filter is for filtering hard solids or large particles and it is not intended for
filtering the water from bacterial contamination.
The filter must be replaced every twelve months at a minimum.
NOTE:
The filter should be replaced only by Belmont Medical
Technologies authorized personnel / authorized biomedical
personnel. See the Service Manual for replacement instructions.