DDT-320-000 Rev G Belmont Medical Technologies
Page
92
of
114
Figure 48: CritiCool MINI System Message Troubleshooting Guide
Message
Cause of Problem
Action to
be taken
Comments
Patient’s temperature is below
XX.X°C
Core temperature is
below the alarm limit
preconfigured in the
settings panel.
The alarm and message
are issued according to
the selected alarm limit.
The available values
are: 31°C, 32°C, 33°C,
34°C, 35°C, 36°C.
Check that the
core temperature
probe is in place
and follow the
patient’s
temperature.
Inform the
physician.
Thermoregulation
continues.
The alarm can be muted for
30 minutes.
Body temperature in accepted
range
Core temperature
reaches an acceptable
core temperature.
The message appears for 5
seconds.
Thermoregulation is Continuing..
CritiCool MINI has left
an alarm state and
returned to a normal
operation mode.
Confirm patient’s
temperature.
The message appears for 5
seconds.
Out of Normothermia Range
Set Point Temperature
for Normothermia is
<32°C and >38.0°C.
For this message,
thermoregulation
continues.
Touch OK to
confirm the new
set point
temperature and
to eliminate the
message.
No alarm.