Operating the System
To turn on the system:
1. Hold down the main power switch at the lower right front of the unit. (See
Front View
on page 18). The Self-Test panel is displayed. At the end of the
Self-Test, the alarm is automatically activated.
Figure 5: Self-Test Screen
2. Following a short Self-Test, the system automatically starts to cool the
water through internal circulation (as in Standby Mode) (See
Figure 10:
Standby
on page 42).
3. Select the appropriate wrap, take it out of the package and place it on
the bed or underneath the patient. (See
Table 2: CureWraps
).
CAUTION!
Power interrupt
The CritiCool
®
MINI has a battery backup that, if charged, will
continue operation for up to an hour if not connected to direct
power. If power does not return within this period, the system
will shut down.
The return of power after shutdown will activate the system
to default settings, irrespective of the mode prior to shut
down.
NOTE:
When using CritiCool MINI in TTM mode, it is highly
recommended to let the CritiCool
®
MINI run before
connecting temperature probes and hoses to allow the
water to cool.
NOTE:
Do not wrap the patient at this time. The Wrap should not
be fastened around the patient until it has filled with water.