9. MAINTENANCE AND TROUBLESHOOTING
Page 9/5
075D0002-01
G - Change contrast
: The display contrast can be adjusted for a better readability. Move the
cursor to the function
7
CHG CONTRAST and press
ENTER
. The CHG CONTRAST menu is
displayed:
CONTRAST
+
:
PRESS ENTER TO VALID
HH:MM
-
:
Press the
UP
key of the numeric keyboard to increase the contrast, or press the
DOWN
key to
reduce the contrast. When the best contrast is obtained, press
ENTER
to validate the setup.
NOTE:
The contrast adjustment menu can be accessed from any other menu by pressing
simultaneously the DEL and "." keys.
H - Park
: Raises the vacuum/waste syringe piston in the upper position and the dilution block
syringes in the lower position.
I - Piercing
: Allows the operator to check the needle height adjustment acccording to the
sample tube holder position. Close the sample compartment door. Move the cursor to the function
9
PIERCING and press
ENTER
: A piercing test is carried out and the needle low position
measured. The following menu is displayed:
If the needle height has to be readjusted, call your local technical support provider or distributor
.
NEEDLE X
CURRENT XXX
STANDARD XXX
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...