8. INSTRUMENT CONFIGURATION
Page 8/5
075D0002-01
Move the cursor next to the required parameter and press
ENTER
, the following menu (RBC for
example) is displayed:
Enter the required low value and press
ENTER
or press the
ESC
key to keep the current value.
Move the cursor to the next required parameter and follow the same procedure. Press the
ESC
key when all modifications have been carried out.
8.2.2. Result high limits
From the CHG LAB LIMITS menu, move the cursor to function
2
HIGH LIMITS and press
ENTER
.
Follow the same procedure for the result low limits.
8.2.3. Print limits and flag values
Diag. 8.1
Move the cursor to function
3
PRINT LIMITS and press
ENTER
. The high and low
limits as well as the flag
values are printed out:
HIGH LIMITS
> 1 WBC HIGH < XX >
HH:MM
2 RBC HIGH
< XX >
RBC LOW ?:
EXIT : ESC
CURRENT :
SAVE : ENTER
NOTE:
PCT and PDW are not available in the United States.
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...