9. MAINTENANCE AND TROUBLESHOOTING
Page 9/9
075D0002-01
Diag. 9.1
Diluent dispenser
: Do you see bubbles in the diluent
dispenser? Is the plunger of the diluent dispenser
moving up and down smoothly during sample
analysis?
Diag. 9.2
If you do not see bubbles and the diluent dispenser
is operating properly, continue. If you see bubbles, or
if the diluent dispenser is not operating properly, refer
to Mechanical Checks for appropriate actions.
If it is, continue. If sample probe is not in its correct
position in the mixing chamber, call your local
technical support provider or distributor.
Diag. 9.3
Transfer of lyse to WBC chamber
: Is the lyse pump
plunger moving up and down smoothly? Can you see
air bubbles?
If the lyse pump is operating properly, and no bubbles
are seen (thus system operations all appear to be
acceptable), call your local technical support provider
or distributor. If bubbles are seen, change the reagent
pack and prime the reagent lines.
Call your local technical support provider or distributor if this does not correct the WBC and HGB
results.
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...