User Manual Closed Tube Model
Daily Startup and Shutdown Procedures
Start-up procedure
1. Press
the
ON/OFF
switch located on the ADVIA
®
60 Hematology System rear panel. Wait for approximately
three (3) minutes.
2. Verify that the printer is connected and turned to
ON
.
3. Wait for the end of the STARTUP CYCLE or press the
STARTUP
key and validate.
4. The background values must be less than:
WBC
0.3 x 10
3
WBC/mm
3
RBC
0.02 x 10
6
RBC/mm
3
HGB
0.0 g/dL
PLT
10 x 10
3
PLT/mm
3
If the background values are not within these limits, the ADVIA 60 Hematology System will automatically
perform another startup cycle. If the STARTUP fails, refer to Section 9 for troubleshooting procedures.
5. Run a blood sample to prime the system:
a. Enter the patient identification or run # (according to the identification mode chosen) using the
ID/SEQ
key.
b. Place the sample tube into the tube holder.
c. Close the door of the tube holder in its sampling position to start the analysis if this starting mode has been
setup, or press the
START
key after closing the door. The tube holder carriage moves up in the piercing
position, and the sample aspiration begins. The green/red LED will blink during sampling.
d. The LED stops blinking and the tube holder carriage moves down when the sampling is completed.
6. Perform a QC procedure using ADVIA 60 TESTpoint
™
Hematology Controls (Prod. No. B03-4200-54, B03-
4201-54, and B03-4202-54).
7. Perform a calibration procedure only if necessary.
8. Run patient samples :
a. Enter patient identification:
•
Press
the
ID/SEQ
key.
•
Enter the identification # (up to 13 characters) or Run # (ranging from 1 to 10000).
•
Press
the
ENTER
key.
b. Place the sample tube into the tube holder.
c. Close the door of the tube holder in its sampling position to start the analysis if this starting mode has been
setup, or press the
START
key after closing the door. The tube holder carriage moves up in the piercing
position, and the sample aspiration begins. The green/red LED will blink during sampling.
d. The LED stops blinking and the tube holder carriage moves down when the sampling is completed.
Shutdown procedure
1. Perform a general cleaning/rinse cycle of the ADVIA 60 Hematology System using the
STAND BY
key on the
front panel. This cycle lasts for approximately one (1) minute. The instrument will cycle into the STAND BY
mode.
2. Press the ON/OFF switch to
OFF
.
NOTE:
When the instrument is left in STAND BY mode, a STARTUP CYCLE must be performed prior to any
analysis cycle
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...