9. MAINTENANCE AND TROUBLESHOOTING
Page 9/16
075D0002-01
N - "BAD VALUE...MINI : XXX, MAXI : XXX"
: this message is displayed in the following
occasions :
- When the sample run number is above 9999.
- When the target values entered in the AUTOCALIBRATION are out of the limits.
- When the number of sample selected in the AUTOCALIBRATION is out of the limits (3 to 11).
- When the calibration factors entered in the CALIBRATION function are out of the factor limits.
- When the laboratory limit values entered in the CHANGE LAB. LIMITS are out of the limits.
- When the flag values entered in the FLAG ADJUST function are out of the limits.
- When the automatic cleaning frequency value entered in the AUTO CLEANING function is out of
the limits.
K - "ERROR : BAD DATE ! TRY AGAIN"
: This message is displayed in the date change function
when the new date entered is incompatible with the date format previously entered. Re-enter a
correct date or change the date format.
L - "ERROR : BAD TIME ! TRY AGAIN"
: This message is displayed in the change time function
when the new time has not the required format : HH.MM. Re-enter the correct time.
M - "CYCLE ABORTED BY USER"
: This message is displayed whenever the
ESC
key is pressed
during an hydraulic cycle. A confirmation message "CYCLE ABORTED?" is displayed. If the
abortion is confirmed by pressing the
ENTER
key, an initialization cycle is carried out to reinitialize
the motors in their home position.
O - "STARTUP NOT INITIATED"
: This message is displayed when the instrument is turned on,
and the startup cycle is not automatically carried out. This STARTUP cycle is mandatory after
each STANDBY cycle in order to flush the instrument from the cleaning reagent. A blank cycle on
reagents is performed to check the cleanliness of the instrument. When this message is displayed
and a sample analysis is requested, press the
STARTUP
cycle key.
P - "STARTUP FAILED, CHECK REAGENTS"
: This message is displayed when the instrument
gives out of range blank values after 3 consecutive startup cycles. Check the expiration dates,
replace the reagents if necessary or perform a concentrated cleaning.
NOTE:
After aborting a hydraulic cycle, it is necessary to run a STARTUP cycle to rinse the
instrument before any further sample analysis.
NOTE:
Maximum and minimum values available are displayed with the corresponding error
message. Correct the values according to the specified indications.
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...