7. CALIBRATION AND QUALITY CONTROL
Page 7/17
075D0002-01
7.5.5. QC statistics
The commercial control files can be printed for permanent records at any time. It is recommended
to print the files at the end of each month.
Each file printout includes the following information: File Name (blood level), Lot number of control,
expiration date of control, date and time of print request, date and time of the run, operator and the
data points for each stored quality control run, the reference means and upper and lower limits,
the actual mean results of the quality controls runs, the 2 standard deviation value, and the
percent coefficient of variation.
7.5.5.1. Select statistics
From the QC MENU, move the cursor to the function
4
STATISTICS and press
ENTER
.
The QC card is read.
7.5.5.2. Select level
Using the
UP
or
DOWN
arrows select the commercial control level to print or select ALL to print
all three levels automatically. Press
ENTER
and the following message appears:
All statistics will be printed.
SELECT LEVEL
3 NORMAL BLOOD
HH:MM
4 HIGH BLOOD
QC
4 STATISTICS
HH:MM
5 GRAPHS
SELECT LEVEL
> * 1 ALL
HH:MM
2 LOW BLOOD
PRINTING QC RESULTS
PLEASE WAIT...
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...