2. INSTALLATION
Page 2/3
075D0002-01
Diag.2.3
2.5. VISUAL CHECKS
2.5.1. Mechanical check
Unscrew the 5 cover fixation screws and remove the
cover: pull it backward and lift it up to the rear of the
instrument.
Diag.2.2
Using the key from the installation kit, unloosen the
locker. Open the pneumatic protection door.
2.4. WORKING CONDITIONS
2.4.1. Environment
ADVIA 60-
CT
should be operated in an indoor location only. Operation at an altitude over 2000
meters (6562 feet) is not recommended. The instrument is designed to be safe for transient
voltages according to INSTALLATION CATEGORY II and POLLUTION DEGREE 2. Please ask
your local technical support provider or distributor for any information about operating location
when it does not comply with the specifications.
2.4.2. Location
ADVIA 60-
CT
should be placed on a clean and level table or work station. Please note that
ADVIA 60-
CT
,
printer and reagents weigh approximately 30 kilograms (66 lbs). Avoid exposure to sunlight.
Proper ventilation requires that a space of at least 20 cm (8 inches) must be left behind the
apparatus.
2.4.3. Grounding
Proper grounding is required. Check that the wall ground (earth) plug is correctly connected to the
laboratory grounding electricity installation. If there is no ground then use a ground stake. Current
electricity norms must be applied.
2.4.4. Humidity and temperature conditions
ADVIA 60-
CT
can function between 18 to 32°C (65 to 90°F). Maximum relative humidity is 80% for
temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C. If it is kept at a
temperature less than 10°C (50°F), the instrument should be allowed to sit for an hour at the
correct room temperature before use.
WARNING: Do not open or close the
instrument front door when the door
of the piercing mechanism is open.
Summary of Contents for ADVIA 60
Page 1: ...075D0002 01 Rev A 2003 09 ADVIA 60 Hematology System Operator s Guide Closed Tube Model...
Page 4: ......
Page 44: ...5 DESCRIPTION Page 5 6 075D0002 01...
Page 74: ...7 CALIBRATION AND QUALITY CONTROL Page 7 20 075D0002 01...
Page 112: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 20 075D0002 01 9 6 PNEUMATIC DIAGRAM...
Page 113: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 21 075D0002 01...
Page 114: ...9 MAINTENANCE AND TROUBLESHOOTING Page 9 22 075D0002 01...
Page 122: ...APPENDIX A SYSTEM AND REAGENT SYMBOLS Page A 4 075D0002 01...
Page 126: ...APPENDIX C BIBLIOGRAPHY Page C 2 075D0002 01...
Page 128: ...INDEX Index 2 075D0002 01...