Field Service Manual
December 2003
Page 200 of 276
ADCPOT on MACCON
Connect the multimeter to the drum diode and set up: on DC, 20 uA.
Execute the following commands at the LTL prompt:
LTL > 0 1 ADCPOT
=>Sets potentiometer in feedback loop of amplifier ADC
=>from Maccon 2.5 the pot meter is always activated. So ‘0 1 adcpot’ is
not needed anymore.
LTL > SHINIT
LTL > SHOPEN
LTL > 4 ONEON
=>switches on beam 4
LTL > 4 XXX SAMP
=> switch on beam 4 with a certain (0 < XXX< 4095) amplification
Choose XXX until you measure 0,35 µA on the voltmeter.
Connect the diode to the ADC input on MACCON.
Then adjust the range of the ADC on MACCON so that the 0.35 µA DC
matches the value 2000
(LTL>rdadc .)
Remark: The higher the resolution the lower the laserintensity has to be on the
film. When high resolutions are used (from 16000ppi and higher) the
ADC on Maccon has to be adjusted to a lower value so it matches with
the value 2000(
➪
LTL>rdadc .) Otherwise the amplitude values of the
DRV4x3 boards will be too low for low laserintensity(par91).
Ex.: 0.2 µA DC matches the value 2000
LTL > DO RDADC . 1E6 WAIT TILKEY
Turn into potentiometer on MACCON till you reach the value of 2000
using the LTL-command RDADC.
Check for all the beams with:
LTL>SHINIT SHOPEN
LTL>2000 0 TNB
Summary of Contents for SilverWriter 800
Page 6: ...Field Service Manual December 2003 Page 6 of 276 ...
Page 8: ...Field Service Manual December 2003 Page 8 of 276 ...
Page 18: ...Field Service Manual December 2003 Page 18 of 276 ...
Page 20: ...Field Service Manual December 2003 Page 20 of 276 ...
Page 24: ...Field Service Manual December 2003 Page 24 of 276 ...
Page 95: ...Field Service Manual December 2003 Page 95 of 276 Multi FilmFeeder solid state panel ...
Page 96: ...Field Service Manual December 2003 Page 96 of 276 ...
Page 122: ...Field Service Manual December 2003 Page 122 of 276 ...
Page 130: ...Field Service Manual December 2003 Page 130 of 276 ...
Page 134: ...Field Service Manual December 2003 Page 134 of 276 ...
Page 168: ......
Page 172: ...Field Service Manual December 2003 Page 172 of 276 ...
Page 174: ...Field Service Manual December 2003 Page 174 of 276 ...
Page 187: ...Field Service Manual December 2003 Page 187 of 276 ...
Page 216: ...Field Service Manual December 2003 Page 216 of 276 ...
Page 220: ...Field Service Manual December 2003 Page 220 of 276 Main directory structure ...
Page 225: ...Field Service Manual December 2003 Page 225 of 276 Adding a DPF file in the plotter queue ...
Page 227: ...Field Service Manual December 2003 Page 227 of 276 Stop logging ...
Page 234: ...Field Service Manual December 2003 Page 234 of 276 ...
Page 240: ...Field Service Manual December 2003 Page 240 of 276 ...
Page 242: ...Field Service Manual December 2003 Page 242 of 276 ...
Page 246: ...Field Service Manual December 2003 Page 246 of 276 ...
Page 248: ...Field Service Manual December 2003 Page 248 of 276 ...
Page 252: ...Field Service Manual December 2003 Page 252 of 276 ...
Page 268: ...Field Service Manual December 2003 Page 268 of 276 ...
Page 270: ...Field Service Manual December 2003 Page 270 of 276 ...
Page 274: ...Field Service Manual December 2003 Page 274 of 276 ...