Field Service Manual
December 2003
Page 104 of 276
When for the first time parameters are put inside the micromaster start with setting p010 to
1(quick commissioning).
And fill in the following parameters.
At the end don’t forget to put parameter 3900 to 3. At that moment all motorparameters are
calculated.
P3900(=3) calculates some parameters depending on motordata, voltage, curent, speed…
Drum Speed Stability :
Drum speed stability seems to be a issue when umaster is cold
Solution:
Par 1910 should be put on 1 and then the motor should be started with belt
disconnected ! !
A message appears A…
The busy sign will shown on the umaster.
After some time the normal display reappears and you can start using the umaster.
P1910 is for measuring different motor parameters.
Parameters Value Comments
R001
read
only
P003
3
3 = expert( See all
params)
P005
21
21= freq / 27=current
P010 0
1=quick
commissioning
P100
0
0 50 Hz operation
2 60 Hz operation USA !!
P300 1
Asynchronous
motor
P304* 230
Volt
P305* 3.6
Ampere
P307* 0.75
Kwatt
P308* 0.82
Cos
phi
P310* 50
Hz
(motor)
P311* 1410 Rpm
P320 0
Motor
magnetizing
current
P335 0
Motor
cooling
P340
0
1= calculation (once)
P341 0.676
P342 375
P601
0
0 = no sensor ptc
P640 180
Motor
overload
factor
P700
2
1=BOP / 2 = digital input
BOP = Front operating
panel
P701
2
1= on/off1
2=on/off1 reverse
P702
4
Quick ramp down
P731 52.3
inverter
running
P844
1
warning motor current
Summary of Contents for SilverWriter 800
Page 6: ...Field Service Manual December 2003 Page 6 of 276 ...
Page 8: ...Field Service Manual December 2003 Page 8 of 276 ...
Page 18: ...Field Service Manual December 2003 Page 18 of 276 ...
Page 20: ...Field Service Manual December 2003 Page 20 of 276 ...
Page 24: ...Field Service Manual December 2003 Page 24 of 276 ...
Page 95: ...Field Service Manual December 2003 Page 95 of 276 Multi FilmFeeder solid state panel ...
Page 96: ...Field Service Manual December 2003 Page 96 of 276 ...
Page 122: ...Field Service Manual December 2003 Page 122 of 276 ...
Page 130: ...Field Service Manual December 2003 Page 130 of 276 ...
Page 134: ...Field Service Manual December 2003 Page 134 of 276 ...
Page 168: ......
Page 172: ...Field Service Manual December 2003 Page 172 of 276 ...
Page 174: ...Field Service Manual December 2003 Page 174 of 276 ...
Page 187: ...Field Service Manual December 2003 Page 187 of 276 ...
Page 216: ...Field Service Manual December 2003 Page 216 of 276 ...
Page 220: ...Field Service Manual December 2003 Page 220 of 276 Main directory structure ...
Page 225: ...Field Service Manual December 2003 Page 225 of 276 Adding a DPF file in the plotter queue ...
Page 227: ...Field Service Manual December 2003 Page 227 of 276 Stop logging ...
Page 234: ...Field Service Manual December 2003 Page 234 of 276 ...
Page 240: ...Field Service Manual December 2003 Page 240 of 276 ...
Page 242: ...Field Service Manual December 2003 Page 242 of 276 ...
Page 246: ...Field Service Manual December 2003 Page 246 of 276 ...
Page 248: ...Field Service Manual December 2003 Page 248 of 276 ...
Page 252: ...Field Service Manual December 2003 Page 252 of 276 ...
Page 268: ...Field Service Manual December 2003 Page 268 of 276 ...
Page 270: ...Field Service Manual December 2003 Page 270 of 276 ...
Page 274: ...Field Service Manual December 2003 Page 274 of 276 ...