6. Optical spares and servicing procedures
Image 6-26
5. Mount the lamp cover and the lamp
Caution:
If you do not mount the lamp cover before starting the projector, the lamps can explode!
6. Select a full white pattern to be projected
7. Turn the adjustment screw until the light beam is aligned on the DMD. This screw can be reached through a whole in the lamp
cover
Image 6-27
Whole in the lamp cover to reach the adjustment screw of the lightrod
78
R59770097 ID H250 / H500 15/05/2007