Connect the suction tube to the suction connector,
making certain that the arrow points away from
MINOP TREND handle
3
.
¾
Connect the irrigation tube to the irrigation con-
nector, making certain that the arrow points
towards MINOP TREND handle
3
.
Note
To facilitate the connection, a MINOP TREND suction/
irrigation tube is available. Its irrigation tube is marked
by a continuous, blue line along its entire length. The
suction tube is not marked.
Connecting the MINOP TREND suction/irrigation
sheath to the MINOP TREND handle
Fig. 4
¾
Slide MINOP TREND suction/irrigation sheath
2
with MINOP TREND endoscope
5
into the rear of
the handle until MINOP TREND suction/irrigation
sheath
2
engages, see Fig. 4.
MINOP TREND suction/irrigation sheath
2
is rotat-
able.
¾
Check the irrigation and suction tubes and the
cone seating for any leaks.
¾
Press irrigation button
4
to check that it functions
properly.
Mounting adapter for holding arm RT099R
Fig. 5
Fig. 6
¾
Undo the cap nut for adapter for holding arm
6
.
¾
Insert adapter for holding arm
6
from either side
(left/right) into the borehole until the flange
engages in the appropriate groove, see Fig. 5.
¾
Screw on the cap nut and tighten it until adapter
for holding arm
6
is fixated, see Fig. 6.
The MINOP TREND (
TR
ansnasal
END
oscopy
System) is now ready for being adapted on the
holding arm.