Aesculap Neurosurgery
MINOP TREND (TRansnasal ENDoscopy System)
4
Assembling
Mounting the MINOP TREND endoscope in the
MINOP TREND suction/irrigation sheath
Fig. 2
¾
Axially insert MINOP TREND endoscope
5
in
MINOP TREND suction/irrigation sheath
2
. When
doing this make certain that the lock of MINOP
TREND suction/irrigation sheath
2
is open and the
pin fits into the groove at the bayonet coupling of
MINOP TREND endoscope
5
, see Fig. 2.
¾
Turn the clamping lever of MINOP TREND suction/
irrigation sheath
2
as far as it will go.
MINOP TREND endoscope
5
is now safely locked in
position.
Connecting the tubes and the irrigation button
Fig. 3
¾
Verify that irrigation button
4
does not show any
damage (valve seals, surfaces, etc.)
¾
Screw irrigation button
4
into MINOP TREND han-
dle
3
down to the limit stop, see Fig. 3.
CAUTION
Damage to components caused by
excessive force!
¾
Handle the components with
appropriate care.
WARNING
Restricted functionality of the
MINOP TREND (TRansnasal
ENDoscopy System) if the irrigation
button is not screwed in
completely!
¾
Make certain that irrigation
button 4 is screwed into MINOP
TREND handle 3 down to the
stop.
CAUTION
Restricted functionality of the
MINOP TREND (TRansnasal
ENDoscopy System) due to
incorrect connecting direction of
the tubes!
¾
Note the arrow on the MINOP
TREND handle.
¾
Carry out function checks
before using the product.