11
Mechanical cleaning/disinfecting with manual pre-cleaning
Manual pre-cleaning with brush
MINOP TREND suction/irrigation sheaths
D–W:
Drinking water
RT:
Room Temperature
Stage
I
¾
Fully immerse the product in the disinfecting
solution. Make certain that all accessible surfaces
are moistened.
¾
Clean the product with a suitable cleaning brush
(e.g. PM995200) until all visible residues have been
removed from the surface.
¾
Brush through all surfaces that are not accessible
to visual inspection, e.g. in products with hidden
crevices, lumens or complex geometries, for at
least 5 min or until no more residues can be
removed. Mobilize non-rigid components, e.g. set
screws, links, etc. during cleaning.
¾
After that, thoroughly rinse (at least 5 times) with
the cleaning solution these difficult to access areas
of the product, using a single-use syringe (20 ml).
¾
To avoid the risk of corrosion, do not use a metal
brush or other abrasives that would damage the
surfaces for cleaning the product.
Stage
II
¾
Rinse the product completely (all accessible
surfaces) under running water.
Stage
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I
Disinfecting
cleaning
RT
(cold)
15
2
D–W
BBraun Stabimed; aldehyde phenol
and QAV free; pH = 9
II
Intermediate rinse
RT
(cold)
1
-
D–W
-