10.2.1 Class II 5,000 lb Models
10.3 Resistance Test for FLSC Weigh Bars
Perform these resistance checks to test the FLSC 05 Weigh Bars. With the meter set
on the appropriate ohms scale, check the following values between the color coded
wires on the section of the cord connected to the Weigh Bar or the corresponding pin
out for the connector. See Figure 10.2.
Figure 10.2 Weigh Bar connector
The electrical connections of the FLSC Weigh Bar can be verified by measuring the
resistances listed below:
Table 10.2 Class II dimensions
Width
Weight "D"
DIM "F"
DIM “K”
30.0"
319 lb
2.0"
4.0
32.0"
339 lb
2.0"
4.0
34.0"
346 lb
2.0"
4.0
36.0"
365 lb
2.0"
4.0
37.0"
369 lb
2.0
4.0
38.0"
373 lb
2.0"
4.0
40.0"
396 lb
2.0"
4.0
42.0"
404 lb
2.0"
4.0
Pins
Wires
Reading
Tolerance in Ohms
1 to 2
ORG to RED
_____________
996 to 1004 (1000)
3 to 4
YEL to BRN
_____________
998 to 1002 (1000)
6 to 7
GRN to BLK
_____________
1074 to 1087 (1080)
7 to 8
BLK to BLU
_____________
1047 to 1053 (1050)
1 to 6
ORG to GRN
_____________
785 to 795 (790)
2 to 6
RED to GRN
_____________
785 to 795 (790)
1 to 7
ORG to BLK
_____________
785 to 795 (790)
2 to 7
RED to BLK
_____________
785 to 795 (790)
3 to 8
YEL to BLU
_____________
798 to 802 (800)
4 to 8
BRN to BLU
_____________
798 to 802 (800)
3 to 7
YEL to BLK
_____________
748 to 752 (750)
4 to 7
BRN to BLK
_____________
748 to 752 (750)
All pins or wires to the metal of the Weigh Bar are to measure >500 mega ohms. Pin
9 (shield, bare wire) to all other pins or wires are to measure >500 mega ohms.
Summary of Contents for FLI 225
Page 10: ...10 FLSC 05 w FLI 225 Instrument Service Manual ...
Page 11: ...FLSC 05 w FLI 225 Instrument Service Manual 11 1 FCC and EMC Declarations of compliance ...
Page 70: ...70 FLSC 05 w FLI 225 Instrument Service Manual Chart of ASCII Characters ...
Page 106: ...106 FLSC 05 w FLI 225 Instrument Service Manual 11 5 Exploded View of FLI 225 AWT05 500348 ...
Page 109: ......