FLSC 05 w/FLI 225 Instrument Service Manual
43
9 Calibrate the FLSC 05 with FLI 225
After Replacing the FLJ 100
Calibration information is stored in both the junction box (FLJ 100) and the instrument.
If the FLJ 100 is replaced, you can retrieve the stored calibration information from the
instrument.
1.
After connecting the FLI 225 to the system, power the instrument on.
The following screen is displayed.
Figure 9.8 FLJ 100 Coefficient Mismatch Screen 1
2.
Press the
Yes
softkey to retrieve the stored calibration coefficient information
from the FLI 225.
The parameters are saved and the instrument returns to normal
operating mode.
3.
Press the
ZERO
key if the weight displayed is not zero.
4.
Proceed to one of the calibration tests on pages 38 through 41.
This option is not available for instruments with EU configuration. The instrument will
display dashes and the message, “Calibration Required.” You must recalibrate the
system.
If the screen in Figure 9.8 does not display on power up, the FLJ 100 that is
connected to the system was set to Factory Defaults. Calibration information will
automatically be uploaded from the FLI 225. Proceed to one of the calibration tests.